MBR-304DRMBR-304GRMBR-304D MBR-304GBlood Bank RefrigeratorINSTRUCTION MANUALMBR-304G
9 REFRIGERATOR COMPONENTS 1. Door switch: To detect the door status (open/close). 2. Battery switch: ON-OFF switch for built-in battery for p
10 REFRIGERATOR COMPONENTS Control panel and keypad 1. Alarm buzzer stop key (BUZZER): To silence the audible alarm, press
11 INSTALLATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the fo
12 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the d
13 BEFORE COMMENCING OPERATION Preparation of monitor bottle Prior to the operation of the unit, fill the monitor bottles (upper and lower)
14 BEFORE COMMENCING OPERATION Remote alarm terminal The terminal for remote alarm is installed inside the rear cover of the unit. The alarm
15 START-UP OF UNIT Follow the procedures for the initial and consequent operations of the unit. 1. Connect the power cord to the dedicated out
16 OPERATING INSTRUCTIONS Storing of items Do not disturb the air circulation in the chamber by the stored items. The disturbed air circul
17 OPERATING INSTRUCTIONS Temperature display The unit is set at the factory to obtain the proper temperature automatically (4oC±1oC). There
18 OPERATING INSTRUCTIONS Calibration of temperature sensor The temperature sensors are attached to the monitor bottles. Follow the inst
1 Note: 1. No part of this manual may be reproduced in any form without the expressed written permission of SANYO. 2. The contents of this manual
19 ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in Table 2, and also self diagnostic functions. Table 2
20 ROUTINE MAINTENANCE WARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent
21 ROUTINE MAINTENANCE Replacement of glow starter A glow starter is located upper right side of the cabinet.. 1. Disconnect the power sup
22 TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when switched on • The
23 SPECIFICATIONS Name Blood Bank Refrigerator Model MBR-304D MBR-304G MBR-304DR MBR-304GR External dimensions W600 x D680 x H1835 (mm) Inter
24 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety ch
7FB6P101379001 (9 Sept. 2004) Recycled paper CEM SANYO Electric Biomedical Co., Ltd.10-15, Hongo 3-Chome
2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and proc
3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain w
4 PRECAUTIONS FOR SAFE OPERATION Disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent e
5 PRECAUTIONS FOR SAFE OPERATION Connect the unit to a power source as indicated on the rating label attached to the unit. Use of any othe
6 CAUTIONS FOR USAGE • If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This
7 ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1): 1. Indoor us
8 REFRIGERATOR COMPONENTS 5 6 12 7 9 8 10 1 2 3 4 13 11 14 15 16 171819 2021 MBR-304G