Sanyo MDF-C8V User Manual Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 92
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
19
F16: <Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F16” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F17: <Purpose> Non-volatile memory initialization, model code change
<Operation> Service code should be input in F06 prior to use this mode.
Input F17 and press SET key to display “001”. Change the value
with key and key. Press SET key to store and revert to PV
display.
Model code ‘003’: MDF-C8V
F18: <Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F18” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F19: <Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F19” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F20: <Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F20” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F21: <Purpose> Serial communication ID setting
<Operation> Input F21 and press SET key to display “000” (initial).
Change the value in the range “001”~”255” with key and key.
F22: <Purpose> Serial communication mode setting
<Operation> input F22 and press SET key to display “000” (initial value)
Change the value with key and key.
Press SET key to store the value and revert to PV display.
Control mode (the 3
rd
digit)
0: Local (initial)
1: Remote
Baud rate (the 2
nd
digit)
0: 2400bps (initial)
1: 4800bps
2: 9600bps
Note) You cannot be changed SV if control mode is set in “Remote”.
F23: <Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F23” displayed to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
F24: <Purpose> To control remote alarm output
<Operation> Input F24 and Press SET key to display “000” (initial).
Change to the desired value with key and key.
Press SET key to store the value and revert to PV display.
000: Remote alarm is ON even if the buzzer stops beeping with BZ
key pressed.
001: Remote alarm is OFF when the buzzer stops beeping with BZ
key pressed.
Page view 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 91 92

Comments to this Manuals

No comments