Sanyo MDF-C8V User Manual Page 68

  • Download
  • Add to my manuals
  • Print
  • Page
    / 92
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 67
65
32
TEMPERATURE RECORDER (OPTION)
6. Attach the sensor cover to the right side of the
chamber with 2 screws on the chamber wall (Fig. 4)
7. Seal the access port on the outside with putty as
shown in Fig. 5.
8. Replace the front panel and fix it with 4 screws.
9. Operate the freezer until the chamber temperature
gets to the set temperature. Check the recorded
temperature and chamber temperature displayed on
the control panel. Adjust the zero adjustment volume
on the temperature recorder so that the recorded
temperature can corresponds with the displayed
temperature if they are not compliance each other.
Refer to the instruction manual enclosed with the
recorder.
Fig. 4
Fig. 5
Putty
Page view 67