6. Press the H
2
O
2
Key for 3 seconds. The system check will start.
7. If the system is normal, the following display will appear. (If the system is not normal, refer to Table 2
on page 57.) Select OK and press the ENTER Key to start H
2
O
2
decontamination. Decontamination will
be performed automatically to step 10. The buzzer will sound when decontamination has been completed
(It takes about 1 hour 40 minutes after warming-up).
Note: The outer door will be locked with an electric lock for safety until the decontamination has been
completed.
8. During the H
2
O
2
mist generation, “H2O2 Decon” will flash at the top left corner of the screen.
Note: H
2
O
2
mist generates from the H
2
O
2
Generator and then quickly changes to a gas.
1
H 2 O 2 D ec o n t a m ination
3
4
R e a d y t o S t a rt?
5
6
M E N U
O K
C a n c e l
1
H 2 O 2 D e c o n
UV Resol ve S t e p:4
2
3
4
5
T e m p : 4 5 . 0
6
D o o r : L o c k e d
7
min
90min
1
H 2 O 2 D e c o n t a m ination
3
4
H 2 O 2 D e c o n S y s tem C h e c k
5
6