INSTALLATION
Assembling of the solenoid valve
1. The joint nut on the outlet side of the CO
2
solenoid valve is removed. The valve outlet pipe with the
pipe insulation is inserted into the joint nut, and fastened in the joint. (Fig. 1) Fasten a joint part securely
so that CO
2
gas can not leak out. (Fig. 2).
2. Insulation is fixed with a SK band. (Two places) (Fig. 3).
CAUTION
Check the gas piping before trial operation to avoid gas leakage. The nut marked with
in Fig. 1 is not fixed surely. Tighten it completely by assembling, and enforce a leak
test.
Fig. 1
Valve outlet pipe
Pipe insulation
LCO
2
solenoid valve
Outlet side
Fig. 2
Insulation
Valve outlet pipe
Joint
Joint nut
Fig. 3
SK band
Mounting plate B